Chapter 7: Changing GX-3R (Pro) Parameters • 89
D Mode Setting
Selected
(factory setting):
LIST
and
LONG.BATT
items do appear in Display Mode.
USER ID
and
STATION ID
screens appear in Display Mode if
ID DISPLAY
in the instrument’s Maintenance Mode
is also set to
ON
(factory setting is
OFF
).
Deselected
:
LIST
,
LONG.BATT
,
USER ID
, and
STATION ID
items do not appear in Display Mode.
Cal Reminder
Selected
(factory setting): The GX-3R (Pro) will give an indication at start up if it is due for
calibration. The type of indication will depend on the
Cal Past Due Act
setting.
Deselected
: There will be no indication at start up if the instrument is due for calibration.
Cal Past Due Act
Confirm to Cal
(factory setting): The GX-3R (Pro) will sound an alarm and the user will have to press
and release the AIR button to continue.
Must Calibrate
: The GX-3R (Pro) cannot be used until a calibration has been performed.
Notification Only
: The GX-3R (Pro) will display that a calibration is due but will continue with the
warm up sequence.
Cal Check Gas (GX-3R Pro Only)
4 GAS:
The calibration dates for only the standard 4 sensors (LEL, O
2
, H
2
S, and CO) are used to
determine if a calibration is due.
ALL
(factory setting): The calibration dates for the standard 4 sensors along with the IR CO
2
or super
toxic sensor are used to determine if a calibration is due.
Man Down (GX-3R Pro Only)
Selected
: The Man Down Alarm can be triggered if the instrument detects no motion for the period of
time defined in
Warning 1 Time
,
Warning 2 Time
, and
Alarm Time
.
Deselected
(factory setting): The Man Down Alarm cannot be triggered.
Panic (GX-3R Pro Only)
Selected
(factory setting): A Panic Alarm can be initiated by firmly tapping the instrument twice in a
row.
Deselected
: A Panic Alarm cannot be initiated.
Warning 1 Time (GX-3R Pro Only)
The
Warning 1 Time
is the amount of time will pass between a Man Down detection and the first
preliminary alarm. It can be set in 1 second increments from
10
-
120
seconds. The factory setting is
60
seconds. Alarm Time
Warning 2 Time
Warning 1 Time.