92 • Chapter 7: Changing GX-3R (Pro) Parameters
are
90
seconds (factory setting), and
120
seconds.
Maintenance Mode Sub Tab
ID Display
Selected
: The User ID and Station ID screens appear in startup sequence. If
D MODE SETTING
in
User Mode 1 Sub Tab is also set to
ON
, then the IDs can be changed in Display Mode.
Deselected
(factory setting): The User ID and Station ID screens do not appear in startup sequence and
the IDs cannot be changed in Display Mode.
Disp Suppress
Selected
: Zero suppression menu item appears in User Mode.
Deselected
(factory setting): Zero suppression menu item does not appear in User Mode. Zero
suppression menu item is always available in Maintenance Mode.
It is not normally necessary to have the zero suppression menu item appear in User Mode. Contact RKI
Instruments before turning this setting on.