Connecting the Printer, Optional • 28
Connecting the Printer, Optional
1 . If you want to power the printer with the battery pack, install the battery pack.
2 . If you want to power the printer with AC power, plug the printer AC adapter’s
wall plug into a wall AC socket.
3 . Connect the AC adapter to the printer’s power jack using the provided cord.
4 . Connect one end of the printer’s USB cable into the printer’s USB connection.
Plug the other end of the printer’s USB cable into the USB printer connector on
the left side of the SDM-6000. Be sure to use the USB cable provided with the
printer. The USB cable provided with the SDM-6000 does not have
appropriately sized USB connections for the printer.
NOTE:
If you have any problems setting up the printer, contact Seiko
Instruments, Inc. at http://www.sii.co.jp/sps/eg/contact.html.
Hardware Assembly
The hardware assembly consists of connecting the AC adapter, installing the air filter,
and connecting the exhaust tubing. Perform the following to complete the hardware
assembly:
1 . Place the SDM-6000(s) on a convenient table top near an AC wall socket in a
well ventilated area. A location near a window that can be opened is best so
that the exhaust can be routed to the window.
2 . Connect the AC adapter’s wall plug into a wall AC socket.
3 . Insert the round plug on the end of the AC adapter’s output cable into the
power jack on the back of the SDM-6000.
NOTE:
If you have multiple SDM-6000s connected to each other, be sure to
plug the multi-station AC adapter into the master station.
4 . Install the air filter so that the arrow on the filter that indicates direction of flow
is pointing towards the AIR fitting.
a. If you are installing the air filter on an SDM-6000, put the green tubing
stub that comes with it on the end with the arrow pointing to it and install
the other end of the green tubing stub on the SDM-6000’s AIR fitting.
NOTE:
If you are testing a GX-6000 with an IR CO
2
sensor in the 0-10,000
ppm or 0-5.00 %vol range, be sure to replace the air filter at the AIR
inlet fitting with the CO
2
scrubber to ensure that the CO
2
present in
fresh air is scrubbed out. The arrow on the CO
2
scrubber should be
pointing toward the AIR inlet fitting.