45 • Setting Basic Operational Parameters in the Config Menu
print calibration results but the results can still be printed by pressing the Print button
at the end of a calibration.
Viewing or Changing the Calibration Parameters
Do the following to set the calibration parameters.
1 . Turn on the SDM-6000 system. The Function Tab will be displayed on a single
station system’s screen or on the master station’s screen in a multi-station
system.
2 . Press the Config button. If Password is set to on (factory setting is off), you will
be prompted for a password. Enter the password and press OK to continue.