65 • Setting Detailed Operational Parameters Using the SDM-PC2 Docking
appears.
To return to the main window without saving any changes, click
Cancel
and
click
OK
in the window that appears.
Updating the Expiration Date Settings
1 . Launch the PC Controller Program and connect your SDM-6000(s) as
described in “Launching the PC Controller Program and Connecting an SDM-
6000” on page 60.
2 . Access the SDM Configuration Window as described in “Accessing the SDM
Configuration Window” on page 61.
3 . Click the Bump & Cal 2 Tab.
Figure 28: Bump & Cal 2 Tab
4 . Select the desired setting for the
Calibration Expiration Notice
parameter.
This is the length of time, in days, before a GX-6000 is due for calibration that
the SDM-6000 will give an indication that the GX-6000’s calibration due date is
approaching. So if calibration is due on 12/20/15 and the
Calibration
Expiration Notice
is set to
10
days, the impending-calibration indication will
begin appearing on 12/10/15. If the GX-6000 is connected to the