12 • Sample Draw Combustible Gas Detector

 

8. When the bag is full, unscrew the fixed flow regulator form the cylinder.

9. Close the clamp, then disconnect the sample tubing from the fixed flow regulator.

 

Performing the response test

 

1. Connect the calibration tubing from the gas collection bag to the inlet line at or near 

the INLET fitting.

The sample draw detector’s pump automatically begins pulling the test sample from 
the gas collection bag when you connect the tubing to the inlet line.

2. After approximately one minute, verify that the reading at the monitoring device 

stabilizes within 

 

±

 

 10% of the concentration of the test sample. If the reading is not 

within 

 

±

 

 10% of the test sample, calibrate the sample draw detector as described in the 

Calibration section of this manual.

3. Remove the calibration tubing from the inlet line, then reconnect the inlet line.

 

Quarterly

 

Calibrate the sample draw detector as described in the “Calibration” section on page 16.

 

Troubleshooting

 

The troubleshooting guide describes symptoms, probable causes, and recommended 
action for problems you may encounter with the sample draw combustible gas detector.

 

NOTE:

 

This troubleshooting guide describes sample draw detector problems only. See 

 

the instruction manual for monitoring device if it exhibits any problems.

 

Fail condition

 

Symptoms

 

The sample draw detector’s Fail light is on.

The monitoring device is indicating a reading well below zero.

The monitoring device is indicating a Fail condition.

 

Probable causes

 

The sample draw detector’s flow rate is too low because of an obstructed sample line, 
failed pump, etc.

The sample draw detector is malfunctioning.

The sensor or transmitter wiring is disconnected or misconnected.

 

Recommended action

 

1. At the sample draw detector, set the correct flow rate with the bypass valve or flow 

adjust potentiometer.

2. If you cannot set the correct flow rate, check the sample lines for obstructions or kinks.

3. Verify that the detector and transmitter wiring are correct and secure. The Installation 

section on page 7 describes detector wiring connections.

4. Calibrate the sample draw detector as described in the Calibration section on page 16.

5. If the fail condition continues, replace the sensor as described later in this section.

6. If the fail condition continues, contact RKI Instruments, Inc., for further instruction.

Summary of Contents for 35-3000RKA-LEL

Page 1: ...detector Table 1 Specifications Target Gas Combustible Gas Input Power 24 VDC Construction housing Fiberglass polyester NEMA 4X Dimensions 8 5 in H x 6 5 in W x 4 25 in D Weight 4 5 lbs Sampling Meth...

Page 2: ...re attached to the back of the housing one at each corner Use the mounting feet to install the housing to a vertical surface Use the two conduit hubs on the bottom of the housing to make wiring connec...

Page 3: ...d or clogged Pump The pump is behind the main circuit board near the top of the sample draw detector The pump pulls the gas sample into the sample draw detector The pump operates on 24 VAC which is ge...

Page 4: ...he EXHAUST fitting accepts 1 4 in rigid tubing See the Installation section on page 7 to connect tubing to the EXHAUST fitting Detection System The detection system consists of the combustible gas sen...

Page 5: ...f the transmitter The transmitter is factory wired to the sensor and main circuit board Test points The test points are located on the left side of the transmitter and are labeled TP and TP A 100 mV 5...

Page 6: ...sure switch are factory wired to the circuit board See the Installation on page 7 for all wiring procedures related to the sample draw detector Relay The relay is to the left of the detector terminal...

Page 7: ...ider the following when you select the mounting site Is there enough room to open the housing door and make wiring connections at the bottom of the housing and tubing connections at the right of the h...

Page 8: ...to seal the connection between the tubing and the INLET fitting See Appendix A Parts List for an example of an appropriate metal insert 2 Place the opposite end of the tubing at the sampling area CAUT...

Page 9: ...e Gas Detector 4 20 Feedback Ground WHT GR N BLK GN D 2 4 V _ LEL O2 AM P P AM P 4 2 0 GREEN BLA C K WH ITE RED PUMP ASSY INTERNALLY WIRED 24VDC GND 24V 4 20 RD PUM P PSW TP TP ZERO SPAN GND 24V 4 20...

Page 10: ...of the sample line with your thumb If the flowmeter ball drops to the bottom of the flowmeter the incoming sample line is not leaking 7 Remove your thumb from the sample line verify the flowmeter retu...

Page 11: ...e target gas Preparing for the response test NOTE This procedure describes the RKI calibration kit that includes a gas collection bag A calibration kit that uses a demand flow regulator is also availa...

Page 12: ...ems you may encounter with the sample draw combustible gas detector NOTE This troubleshooting guide describes sample draw detector problems only See the instruction manual for monitoring device if it...

Page 13: ...pump etc The sample draw detector is malfunctioning The demand flow regulator is not operating properly if one is used for calibration instead of a gas bag Recommended action 1 Verify that the calibra...

Page 14: ...row indicates the direction of the sample flow 3 Disconnect the filter from the elbows on each end of the filter then remove the filter from the sample draw detector 4 Make sure the arrow is pointing...

Page 15: ...low to 0 6 SCFH 2 Slowly turn the low flow potentiometer counterclockwise just until the sample draw detector goes into low flow alarm NOTE The low flow potentiometer is accessible through a circular...

Page 16: ...the correct test points output for the calibrating sample Output mV calibrating sample fullscale X 400 100 For example with a calibrating sample of 50 LEL and a fullscale setting of 100 LEL the corre...

Page 17: ...e sample tubing from the inlet line then close the clamp Setting the Response Reading 1 Connect the sample tubing from the gas collection bag to the fixed flow regulator then open the clamp 2 Screw th...

Page 18: ...Filter Boston DFU9933 05 DQ 57 1050RK Transmitter specify target gas when ordering 61 0145RK Combustible gas sensor NC 6240 plug in type LEL range 81 0002RK 01 Calibration cylinder 50 LEL Hydrogen 34...

Reviews: