Maintenance and service
•
With the "Reagent" key, you can dose reagent into the
measuring chamber to produce a colouration of the sample
water
•
Depending on the water hardness and the dosed reagent
quantity, you will get a colour change, e.g., for total hardness
from red to green.
•
The turnover point is not a measure of water hardness because
an undefined amount of reagent is dosed.
•
The "Valve" button can be used to flush the reagent out of the
measuring chamber.
Exit diagnostic program
After checking the unit functions, switch the programme switch
SW4 back to the OFF position = analysis mode.
After the diagnostic programme has been carried out, switch all programme switches
back to the original position or according to the local operating conditions.
Diagnostic functions LIMESBase
The LIMESBase does not have a diagnostic programme.
To unlock the diagnostic programme, you need an upgrade kit, which you can purchase from us.
For more information, see page 8, 21 and 68