English
14 of 20
FreedomEdge® Flow Profile:
The FreedomEdge® flow profile shows that the flow rate is consistent throughout the
delivery of medication.
FreedomEdge
®
Flow Rate vs. Time
Fluid: H
2
0 · Fluid Volume: 20ml · Tubing Measured: F120, nominal 133.55ml/hr
0.00
20.00
40.00
60.00
80.00
100.00
120.00
140.00
160.00
0
75
150
225
300
375
Fl
ow
R
at
e
(m
L/
hr
)
Time (seconds)
Technical Specifications:
System
Reservoir volume:
20ml/30ml
Residual volume:
<0.4ml
Flow rate accuracy:
± 8%*
Operating pressure:
15psi (peak)/13.5psi (nominal)
(20ml reservoir)
Height sensitivity:
± 3% per 12” (30cm)
*
Flow rate data recorded at 21°C using 20ml of 0.9% NS. An overall accuracy of ±8% is expected at
these values. At higher temperatures, which results in a decrease in viscosity, a higher flow rate is
anticipated to occur. The flow rate variation due to changes in temperature is approximately linear
and would vary from -20% at 14°C to +20% at 30°C. Fluids considerably more viscous than 0.9% NS
are expected to slow the flow rate and consequently may result in longer infusion times. More
viscous fluids may be tested before patient use by following the bench test procedures described
in this manual.
Summary of Contents for Freedom Edge