13
Specifications
General Specifications
Display Count :
6000 count.
Overrange display:
“OL”
Conversion Rate :
2 ( times / second ) .
Dimensions (W x H x D):
60mm x 147mm x 31.5mm
Weight:
140g.
Power Requirement:
LR44 Size Button Battery 1.5V x 2.
Maximum Conductor Size:
22mm.
LVD:
EN61010-1, EN61010-2-030, EN61010-2-032
EMC:
EN 61326-1
Installation Category:
CAT.III. 600V
CAT
Application field
I
The circuits not connected to mains.
II
The circuits directly connected to Low-voltage installation.
III
The building installation.
IV
The source of the Low-voltage installation.
Environmental Conditions
Indoor Use.
Pollution degree:
2
Operating altitude:
2000m (6562 ft)
Operating temperature:
0°C ~ 30°C (≤80% RH),
30°C ~ 40°C (≤75% RH),
40°C ~ 50°C (≤45%RH)
Storage temperature:
-20°C to 60°C, 0 to 80% R.H. (Batteries not fitted).
Temperature coefficient:
0.2 x (Spec.Accy) / °C, < 18°C or > 28°C .
Shock vibration:
Sinusoidal vibration per MIL-PRF- 28800F (5 ~ 55 Hz, 3g max.)
Drop Protection:
4 feet drop to hardwood on concrete floor.