14
Electrical Specifications
Accuracy is given as ±(% r number of digits) at 23°C ± 5°C < 80%RH. Accuracy is
specified for a period of one year after calibration.
AC Function
ACV and ACA specifications are ac coupled, true R.M.S.
The crest factor may be up to 3.0 as 4000 counts.
For non-sinusoidal waveforms, Additional Accuracy
by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Position Error of Clamp: ±1.5% of LCD reading.
Range
Resolution
Accuracy
60.00 A(1)
0.01 A
± (1.5% + 10d)(2)
300.0 A
0.1 A
± (1.5% + 5d)
DC Current
Range
Resolution Accuracy(50~100Hz)
Accuracy(100~400Hz)
60.00 A
0.01 A
± (1.5% + 25d)
± (2.5% + 25d)
300.0 A
0.1 A
± (1.5% + 5d)
± (2.5% + 5d)
AC Current
(1) There are less than 0.3A variation as measuring in different directions.
(2) Add 10D to accuracy in Auto AC & DC Sense Mode.
Frequency Response:
50~400Hz (Sine Wave).