29
The option contains identification code (1-9999) for the fixture, which is used for the master/slave operation.
- Adjusting the default positions of the CMY wheels
By this function you can adjust the colour wheels and the dimmer unit to their right positions. Use the [Up] and
[Down] buttons to browse through the adjusting menu - the display shows step by step these messages: "CYAn,
MAGE,YELL,dimr" by which you can adjust the effects to the required positions (0-255) before the function
calibration.Then when the positioning is finished use the last "FCAL" function (Fixture calibration).
1. Calibration via the control board
Press [Enter] and the [Up] and [Down] buttons in order to display the following messages: "CYAn, MAGE,YELL,"
for very smooth function calibration. Select one of them, press [Enter] and use the [Up] and [Down] buttons in
order to adjust their right value from 0 to 255. Then press [Enter] to confirm or [Mode] to cancel and return to the
menu. This can be repeated for each calibration parameter if it is required. When the calibration is finished, it is
necessary to use the "A.rES." function in order to write the calibration values to the memory (EPROM) and to
make a reset in order to check the newly adjusted positions of the CMY-colours. When the reset of the fixture is
finished, the display will show the "F.CAL." message. Press [Enter] to repeat the calibration or [Mode] to return
to the "AdJ" menu.
2. Calibration via the external controller
Connect the DMX controller to the fixture.Press [Enter] and the [Up] and [Down] buttons in order to display the
following messages: "CYAn, MAGE,YELL" - calibration parameters. Select one of them and press [Enter].
Now you can calibrate the CMY-colours by your controller. The DMX calibration protocol is described in the table
mentioned below.
DMX Calibration protocol:
After having calibrated required functions press [Enter] to confirm (or [Mode] to cancel and return to the menu
without reset by the "A.rES." function) and use the "A.rES." function in order to write the calibration values to the
memory (EEPROM) and to make a reset in order to check the new adjusted positions of the CMY-colours .
Summary of Contents for ECOLOR 250 XT
Page 1: ......