Glucotrend

®

 
Basic Meter Instructions  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Coding the Glucotrend

®

 

Remove the calibration coding chip from the test strip pack 

 

The coding chip slot is level with the bottom of the display window. 

 

Turn the chip so that the flat side with the code number inscribed 
on it, is facing upwards. 

 

Carefully insert the coding chip in the slot and press it in as far as it 
will go. 

 

If the monitor is off, switch it on. 

 

The bottom line of the display will show the time and date and a 
three digit code will be shown in the centre of the display. To the 
left of the screen, a test strip symbol and CODE will be flashing. A 
red dot flashes in the test-strip guide. 

 

Make sure that code on the monitor is the same as the code 
number on the pack of test strips.  

 

Performing a test 

 

Ensure you have set the time and date on your Meter before 
testing. Please refer to "Entering Settings" in the User's Guide. 

 

Turn the meter on by pressing centre button 

 

To check that all display elements are correct, wait for all three 
digits 888 to completely appear on the screen  

 

Check that the code on the screen matches the code number 
printed on the test strip vial. Remember to change the code 
calibration chip every time you open a new vial of test strips.  

 

Remove test strip from the vial. Replace vial cap. 

 

Insert the strip in the test slot in the direction shown by arrows until 
it clicks into place. 

 

The 3 digit code number will disappear. 

 

After approximately one second, a small blood drop symbol flashes 
on display indicating readiness to apply blood to test strip.  

 

After just a few seconds, you will hear a beep tone and a rotating 
box will indicate that the reading is in progress. Result will show in 
15 seconds.

 

 

Results will be saved automatically, if time and date are set 
correctly.  