27
Control Tests
4
When to Perform a Control Test
Performing a control test lets you know the meter and test strips are working properly. You should perform a
control test when:
• you open a new test strip box.
• you left the test strip container open.
• you think the test strips are damaged.
• you want to check the meter and test strips.
• the test strips were stored in extreme temperatures, humidity, or both.
• you dropped the meter.
• your test result does not match how you feel.
• you want to check if you are performing the test correctly.
About the Control Solution
• Use only Accu-Chek Performa control solution.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3 months from the date the control solution bottle was opened (discard date) or on the use by
date on the bottle label, whichever comes first.
• Do not use control solution that is past the use by or discard date.
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the difference between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.
Chapter 4: Control Tests
51826_07054246001_01_EN.indb 27
9/26/13 2:07 PM
Summary of Contents for Accu-chek Performa
Page 2: ...51826_07054246001_01_CVR indb 2 9 12 13 4 05 PM ...
Page 4: ...51826_07054246001_01_EN indb 2 9 26 13 2 07 PM ...
Page 38: ...36 Maintenance and Troubleshooting 5 51826_07054246001_01_EN indb 36 9 26 13 2 07 PM ...
Page 44: ...42 Technical Information 6 51826_07054246001_01_EN indb 42 9 26 13 2 07 PM ...
Page 46: ...44 Notes 51826_07054246001_01_EN indb 44 9 26 13 2 07 PM ...
Page 47: ...45 Notes 51826_07054246001_01_EN indb 45 9 26 13 2 07 PM ...
Page 48: ...46 Notes 51826_07054246001_01_EN indb 46 9 26 13 2 07 PM ...
Page 49: ...47 Notes 51826_07054246001_01_EN indb 47 9 26 13 2 07 PM ...
Page 50: ...48 Notes 51826_07054246001_01_EN indb 48 9 26 13 2 07 PM ...
Page 52: ...51826_07054246001_01_FR indb 2 9 26 13 2 19 PM ...
Page 86: ...36 Entretien et dépannage 5 51826_07054246001_01_FR indb 36 9 26 13 2 19 PM ...
Page 92: ...42 Informations techniques 6 51826_07054246001_01_FR indb 42 9 26 13 2 19 PM ...
Page 94: ...44 Remarques 51826_07054246001_01_FR indb 44 9 26 13 2 19 PM ...
Page 95: ...45 Remarques 51826_07054246001_01_FR indb 45 9 26 13 2 19 PM ...
Page 96: ...46 Remarques 51826_07054246001_01_FR indb 46 9 26 13 2 19 PM ...
Page 97: ...47 Remarques 51826_07054246001_01_FR indb 47 9 26 13 2 19 PM ...
Page 98: ...48 Remarques 51826_07054246001_01_FR indb 48 9 26 13 2 19 PM ...
Page 100: ...51826_07054246001_01_PT indb 2 9 26 13 2 24 PM ...
Page 134: ...36 Manutenção e diagnóstico de anomalias 5 51826_07054246001_01_PT indb 36 9 26 13 2 24 PM ...
Page 140: ...42 Informações técnicas 6 51826_07054246001_01_PT indb 42 9 26 13 2 24 PM ...
Page 142: ...44 Notas 51826_07054246001_01_PT indb 44 9 26 13 2 24 PM ...
Page 143: ...45 Notas 51826_07054246001_01_PT indb 45 9 26 13 2 24 PM ...
Page 144: ...46 Notas 51826_07054246001_01_PT indb 46 9 26 13 2 24 PM ...
Page 145: ...47 Notas 51826_07054246001_01_PT indb 47 9 26 13 2 24 PM ...
Page 146: ...48 Notas 51826_07054246001_01_PT indb 48 9 26 13 2 24 PM ...
Page 147: ...48 مالحظات 51826_07054246001_01_AR indb 48 19 09 2013 18 20 49 ...
Page 148: ...47 مالحظات 51826_07054246001_01_AR indb 47 19 09 2013 18 20 49 ...
Page 149: ...46 مالحظات 51826_07054246001_01_AR indb 46 19 09 2013 18 20 49 ...
Page 150: ...45 مالحظات 51826_07054246001_01_AR indb 45 19 09 2013 18 20 49 ...
Page 151: ...44 مالحظات 51826_07054246001_01_AR indb 44 19 09 2013 18 20 49 ...
Page 153: ...42 صنفصاوتافصاوملا 6 51826_07054246001_01_AR indb 42 19 09 2013 18 20 49 ...
Page 159: ...36 ةنايصلا نملاوءاةخألاةنالعو 5 51826_07054246001_01_AR indb 36 19 09 2013 18 20 47 ...
Page 193: ...51826_07054246001_01_AR indb 2 19 09 2013 18 20 21 ...