9
en
Picture 8
Attach the luer-lock (K) to the pump by tightening
it into the pump adapter. Follow the pump manu-
facturer’s user manual to prime the transfer-set.
Prime the transfer-set until medication drips
from the connector needle tip. Do not leave air
bubbles in either the cartridge or the transfer-set.
Connect the transfer-set to the newly inserted
head-set and fill the empty space of the cannula
housing with a bolus of 1.0 units (U100 insulin).
WARNING
You must fill the empty space of the newly in-
serted head-set after connecting the transfer-
set. Failure to do so results in missed insulin.
If the coupling does not click noticeably and au-
dibly together, use a new head-set and a new
transfer-set.
Do not use any tools to screw the luer-lock con-
nector into place.
Filling volume transfer-set:
12”/ 30 cm: approx. 6 units* approx. 60 μl
24”/ 60 cm: approx. 11 units* approx. 110 μl
31”/ 80 cm: approx. 14 units* approx. 140 μl
43”/110 cm: approx. 18 units* approx. 180 μl
Filling volume head-set:
approx. 1.0 units* approx. 10 μl
*U100 Insulin