Protection against infection:
r
When collecting samples always observe the
general precautions and guidelines relating to
blood sampling.
u
r
Dispose of all test strips used for patient testing in
accordance with the disposal policy of your
laboratory or practice.
u
u
8.2 Cleaning and disinfecting the meter (155)
Never...
¾
store the meter at high temperatures (above 35 °C
or 95 °F) while it is in regular use.
¾
store the meter in damp or humid conditions
without protection.
¾
remove or insert the code chip while the meter is
performing a test.
¾
touch or remove the test strip during a test.
¾
wait more than 15 seconds after a fingerstick
before applying the blood.
¾
fingerstick a patient with wet hands (residues of
water, sweat, or alcohol).
¾
add more blood after the test has begun.
¾
perform a test with a drop of blood from a previous
puncture.
CAUTION!
Accuracy/precision of measured results:
Failure to comply with the above may lead to
inaccurate results. An incorrect result may lead to an
error in diagnosis, therefore posing danger to the
patient.
4 Testing a Blood Sample
• 4
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......