105
CoaguChek
®
Pro II · Operator’s Manual
4.2 Preparing to test
r
To prepare to test
1
Have the test strip container at hand.
2
Make sure that the code chip supplied with these
test strips is at hand.
Each box of test strips contains a code chip. The
parameter, the number on the code chip and the
number on the test strip container must match. A
capital
S
in front of the number indicates that this
code chip is for test strips. (A capital
C
in front of the
number indicates that it is a control solution code
chip. See Chapter 5, starting with
Control testing
).
Test strip code chip
The code chip provides the meter with important
information that it needs to perform the
coagulation test. The chip contains information
about the test method, the lot number, and the
expiration date. The code chip is required,
whenever a new test strip lot is used, so that the
meter can read and store the lot information about
that particular lot of test strips.
The CoaguChek Pro II meter stores the data from
up to 60 test strip code chips that have been
inserted.
•
Do not forget to use the test strip code chip
that is supplied with each pack of test strips
before you perform the first test with these
strips. We recommend that you leave the code
chip in the meter to protect the electrical
contacts in the meter from becoming dirty.
4 Testing a Blood Sample
• 4
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......