114
CoaguChek
®
Pro II · Operator’s Manual
The hourglass icon shows that the test strip is
warming up. When the warming‑up process is
complete, a further beep (provided the beeper is
enabled) indicates that you can now apply blood.
The blood drop icon flashes to indicate that the
meter is ready to perform the test and is waiting
for blood to be applied.
15
A countdown begins. You must apply the drop of
blood to the test strip before the countdown ends.
Otherwise you will receive an error message.
When the test is running, no operator action is
required. The display dims to save energy.
r
To apply blood from the fingertip
WARNING!
Use only a lancing device that is approved for use
by healthcare professionals, such as the
Accu‑Chek Safe-T-Pro or Safe‑T‑Pro Plus lancing
devices, or the CoaguChek Lancets
(professional‑use lancets available only in the
United States). Follow the manufacturer’s
instructions.
1
Now lance the side of a fingertip with the lancing
device.
We recommend obtaining the capillary blood from
the side of the fingertip as this causes the least pain.
4 Testing a Blood Sample
• 4
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......