117
CoaguChek
®
Pro II · Operator’s Manual
3
Results that are above or below the measuring
range are indicated by the symbols > (above) or <
(below).
•
If a “C” is displayed along with the result:
This may occur if the hematocrit value is very
low or due to erroneous blood collection (e.g.,
wet hands). Repeat the test. If using capillary
blood, make sure that the patient's hands are
dry. If the message persists, perform a
hematocrit check.
•
If a
“
”
is displayed along with the result:
This may occur if the test result is out of
normal range. Touch the
icon to display the
Out Of Normal Range
information screen.
When interpreting results, refer to the detailed
information on limitations and interferences included
in the limitations section of the test strip package
insert.
r
To accept/reject a test result
1
If this function is enabled, you may choose to
accept or reject a test result. When the result is
displayed, touch
•
Reject
or
•
Accept
4 Testing a Blood Sample
• 4
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......