125
CoaguChek
®
Pro II · Operator’s Manual
6
Hold the dropper with the sealed dropper neck
pointing upward, then cut off the end of the cap
with scissors. Do not hold the dropper close to
your face.
To avoid loss of diluent, hold the dropper by the
stem; do not squeeze the bulb of the dropper while
cutting the tip.
7
Apply gentle pressure to the reservoir to transfer
the entire contents of the dropper to the bottle.
Make sure that the dropper does not come into
contact with the dried control plasma.
8
Close the bottle again.
9
Make sure the dropper is at hand for the next
steps in the liquid quality control test.
5 Control and Proficiency Testing
• 5