After the required contact time, carefully dry the
test strip guide cover with a dry lint‑free cloth.
5
NOTICE!
Damage to the instrument
Do not insert any objects into the test strip guide.
Doing so could damage the electrical contacts
behind the test strip guide.
Disinfect the interior test strip guide.
Hold the meter upright with the test strip guide
facing down.
Wipe the easily accessible area (shown in the
illustration: white area and its black surrounding
frame) with a fresh germicidal cloth. Do not clean
beyond the white area inside the meter.
Ensure the germicidal cloth is only damp, not wet.
To properly disinfect, let the test strip guide
sit wet for a contact time of at least
2 minutes, or as instructed by the
manufacturer in the product labeling.
Always refer to the manufacturer’s labeling
for contact time recommendations.
6
Dry the interior test strip guide.
After the required contact time, carefully dry the
test strip guide with a dry lint‑free cloth.
Let the inside of the test strip guide dry for at least
10 minutes before re‑attaching the test strip guide
cover and testing again.
Visually verify that no residual moisture is seen
8 Maintenance and Care
• 8
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......