199
CoaguChek
®
Pro II · Operator’s Manual
infrastructure, DMS workload, etc. Once the DMS
sends a query, however, the meter will respond
within a few seconds.
¾
A CoaguChek Pro II meter with wireless
connectivity enabled, communicates results after
every test. When the meter is idle while docked, it
automatically attempts to communicate with the
DMS every 30 minutes. Whenever the meter is
placed in the HBU or connected to the power
adapter for recharging, it will automatically attempt
to communicate with the DMS if the battery is still
sufficiently charged.
A typical range for direct connection between the
CoaguChek Pro II meter and the access point (air,
direct view, low disturbances) is up to 15 to 20 meters
(49 to 66 feet). The actual range depends on the
positioning of the access point’s antennas and other
topological properties of the space between WLAN
device and AP. Additionally, dynamic control of the
transmitting power of the access point may reduce the
maximum distance between WLAN device and AP
within which communication can be guaranteed.
The CoaguChek Pro II system is designed such that it
coexists with other wirelessly communicating devices.
The CoaguChek Pro II system does NOT include any
real-time or even time critical wireless functionality. It
communicates exclusively single, digital data fields. It
does NOT communicate continuous waveform data.
Note: A degraded WLAN connectivity will not impact
the functionality of the meter but may delay the
communication of results to the DMS. Users should
be aware that
real‑time communication
of data
cannot be guaranteed
by the CoaguChek Pro II
meter.
During the exchange of large data packages a
short‑term delay in the meter response might occur.
14 Appendix C
• 14
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......