50
CoaguChek
®
Pro II · Operator’s Manual
Group
Subgroup
Setting
Values
(a)
Off
Delete Database
Results
Code Chip Data
Entire Database
Parameter Selection
Enable
Disable *
ID Setup
Admin.
(Administrator)
Blank (Off) *
Operator (Operator List is optional)
Inactive *
Active
Scan Only/Hidden list
Patient
No *
Optional
Required
Scan Only/Hidden list
QC Settings
QC Range
PT
Default Range *
aPTT
Custom Range
Lockouts
Operator Lockout
(only if the Operator
option is set to Active
and an operator list is
available)
Off *
Weekly
Monthly
Every 3 months
Every 6 months
Yearly
Number of levels
QC Lockout
New code (Yes/No)
Scheduled (No*,
Daily, Weekly,
Monthly)
Number of hours
(a)
Default settings are labeled with an asterisk (*).
3 Meter Setup
• 3
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......