95
CoaguChek
®
Pro II · Operator’s Manual
The setting for
Time of day
applies to all lockouts
based on date intervals. The setting for
First day of
week
applies to lockouts in weekly intervals only.
6
Touch
or
to enter the desired time and day.
7
Touch
to save this setting, or touch
to exit
this menu without saving any changes. The display
automatically returns to the previous screen.
STAT Test Configuration
Quality control tests ensure consistent quality and
accurate measurements. In emergency situations,
however, it may be necessary to perform a test without
delay. To override an active lockout
(3)
, you can allow
STAT Tests (STAT =
S
hort
T
urn
A
round
T
ime).
You can set the number of tests (up to nine) permitted
beyond the lockout. Once the number of STAT Tests
allowed has been reached, additional tests cannot be
performed until you have successfully completed a
quality control test. The default setting is
three
STAT
tests.
The results of STAT tests are always identified by the
label
STAT Test
when they are displayed.
u
(3)
QC, operator, or new lot lockout
3 Meter Setup
• 3
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......