110
CoaguChek
®
Pro II · Operator’s Manual
4.3 Performing a test
r
To perform a test
1
Check the battery level.
•
If the battery icon turns red (
one bar
remaining
), there may not be enough power
remaining for another test.
•
If there are
no bars
remaining in the battery
icon, you cannot perform any more tests.
Power the meter off by pressing
.
In both cases, restore power by recharging the
battery pack, or using the power adapter.
2
Check that the date and time are correct. Correct
any wrong entries as described in Chapter 3,
Meter Setup
.
3
If a lockout (
Operator
or
QC Lockout
) is displayed
instead of the
Patient Test
button, you must run a
liquid quality control test before you can perform a
patient test. See Chapter 5,
Control and Proficiency
Testing
, starting with
When the meter is in lockout status, a patient test
cannot be performed.
•
The next steps depend on whether you work
with
or
without
patient lists.
4
If the
Patient Test
button is available, but a
lockout is displayed (QC!), one or more test
parameters are locked. A test can only be
performed for parameters that are still unlocked or
4 Testing a Blood Sample
• 4
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......