128
CoaguChek
®
Pro II · Operator’s Manual
6
Touch
Control Test
.
7
In the
QC Test
menu, touch
Control Test
again.
8
Touch the button to select the parameter you want
to use.
9
The test strip icon prompts you to insert a test
strip. Remove a test strip from its container and
close the container again with the stopper.
Exposure to external influences (such as humidity)
may deteriorate the test strips and may lead to error
messages. Therefore, always close the test strip
container immediately after removing a test strip.
10
Hold the test strip so the lettering with the test
name is facing upward.
11
Insert the test strip into the test strip guide in the
direction indicated by the arrows. Slide the test
strip in as far as it will go.
A beep indicates that the meter has detected the
test strip (provided the beeper is enabled).
5 Control and Proficiency Testing
• 5
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......