129
CoaguChek
®
Pro II · Operator’s Manual
12
If you are using a new test strip lot and have not
inserted the test strip code chip yet, you must do
so now. Otherwise you cannot perform a quality
control test.
As with the test strips, a quality control code chip
is also provided with the control solutions. This
chip informs the meter about the acceptable
ranges of results for that lot of controls. The
information on the code chip is retained in the
memory so you can use the same control solutions
at any time.
13
Select the code stored for your current control
solution or touch
New Code
to use a new control
solution.
When entering or selecting a code manually, make
sure it matches the code printed on the bottle of
the control solution.
14
Alternatively, the control solution code can also be
entered
using the built‑in barcode scanner
.
Touch
Scan
and scan the barcode on the label
from a distance of approx. 10 cm (4 inches).
The meter beeps once the barcode has been read
successfully. The barcode information appears in
the code field. The scanner turns off after
10 seconds, if a barcode is not scanned.
5 Control and Proficiency Testing
• 5
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......