Using the dropper, draw up the dissolved contents
of the bottle.
18
Apply a single drop of control solution directly
from the dropper to the semicircular, transparent
sample application area on top of the test strip. Do
not add more control solution.
You hear a beep when you have applied enough
control solution (provided the beeper is enabled).
The dropper icon disappears and the test starts.
The result of the liquid quality control test is
displayed. It is automatically saved to memory.
The acceptable range of results for the liquid
control is displayed below the current result, along
with Pass or Fail.
If a quality control test fails, an up arrow (too high)
or down arrow (too low) is displayed and flashes.
5 Control and Proficiency Testing
• 5
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......