148
CoaguChek
®
Pro II · Operator’s Manual
The option of setting up an
Operator Lockout
is
available
only
when operator lists are created on the
DMS, stored in the meter, and Operator login is
activated. See
Operator login
and
Operator Lockout
for more details.
u
u
Computer (Setup option)
For initial connection to a DMS, the ability to
communicate within a network has to be set up as
follows:
¾
In the
Setup Menu
, the
Connection
option must be
set to
Computer
.
¾
The Handheld Base Unit must be configured
correctly. (For details please refer to the manual of
the Handheld Base Unit and the Technical Note
stored on the Handheld Base Unit itself.)
¾
Place the meter on a Handheld Base Unit that is
connected to the network. The meter will
automatically be recognized by the DMS and may
now be assigned to a site and/or unit managed by
the DMS (if your DMS has that functionality).
¾
Depending on DMS functionality available, the
DMS may transfer settings and lists to the meter
as well as enable test results to be transferred
from the meter to the DMS.
u
To enable the connection option (66)
Operator lists
After powering the meter on, an
Operator
login may be
configured. There are four possible configurations for
an
Operator
login:
¾
Operator
is activated, but there is no list available:
The
Operator
can either log in via the onscreen
keypad or a barcode scan. The use of the meter is
not restricted to a predefined
Operator
group, so
Operator ID
s are entered via the onscreen keypad
or read from the barcode and stored with test
results for informational purposes only.
7 Extended Functionalities
• 7
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......