156
CoaguChek
®
Pro II · Operator’s Manual
Follow recommendations from the FDA, CDC and
CMC and your facility’s policies and procedures for
infection control.
(9),(10)
The FDA recommends that Point of Care testing
devices should be used on one patient only and not
shared. If it is not possible for each individual patient
to have a dedicated POC device, the meters must be
properly cleaned and disinfected after every use or
before being used on the next patient, following the
guidelines given below.
(11)
Approved cleaning and
disinfecting products
The following active ingredients are tested and
approved for cleaning and disinfecting the meter
housing and test strip guide:
We recommend Super Sani‑Cloth
®
Germicidal
Disposable Wipes (disposable wipes containing 0,5 %
quaternary ammonium compounds [single compound
or mixture] and up to 60% isopropyl alcohol
[isopropanol]). This product contains the approved
active ingredients and is licensed for use in Canada.
The use of Sani‑Cloth
®
Plus Germicidal Disposable
Wipes is not recommended.
NOTICE!
Damage to the instrument
r
Use only products containing the approved active
ingredients. Do not use any other cleaning or
disinfecting solutions as this could result in
damage to the system components.
(9)
FDA Public Health Notification: Use of Fingerstick Devices on More than One Person Poses Risk for
Transmitting Bloodborne Pathogens; Initial Communication, (2010). http://www.fda.gov/
MedicalDevicesSafety/ AlertsandNotices/ucm224025.htm
(10)
CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for
Transmitting Bloodborne Pathogens, (2010). http://www.cdc.gov/injectionsafety/
FingerstickDevicesBGM.html
(11)
Healthcare Infection Control Practices Advisory Committee (HICPAC), William A. Rutala, PhD., M.PH.,
and David J. Weber, M.D., M.PH. Centers for Disease Control and Prevention, 2008. Guideline for
Disinfection and Sterilization in Healthcare Facilities. Atlanta, GA
8 Maintenance and Care
• 8
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......