165
CoaguChek
®
Pro II · Operator’s Manual
8.5 Maintenance cleaning of the interior test strip guide in case of residue
build‑up
When to clean the test strip guide
Over time, the Super Sani‑Cloth
®
Germicidal
Disposable Wipes may leave a build‑up of residue in
the test strip guide.
¾
The test strip guide should be visually inspected
and cleaned periodically as necessary or at least
once per month.
Approved cleaning agents
Test strip guide: Use a cotton swab slightly dampened
with water.
NOTICE!
r
Do not use any other cleaning solutions as this
could result in damage to the system components.
r
To perform maintenance cleaning of the
test strip guide
1
Open the cover.
With the meter powered off, use your thumbnail to
open the cover of the test strip guide by pressing
its front edge upward. Move the cover safely away
from the meter.
Clean the inner side of the cover with a cotton
swab and water. If necessary rinse the cover under
tap water.
8 Maintenance and Care
• 8
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......