CoaguChek
®
Pro II · Operator’s Manual
meter, and Operator login is activated. These lists are
only available in connection with a data management
system. For more details see
Data handling
.
u
2 Level QC
With this option you can configure a continuous
2‑level liquid quality control test, if the test has two
levels. The second test will start immediately after
successfully finishing the test for one level. If disabled,
the meter returns to the
Main Menu
once one of the
two levels has successfully been tested.
Scan QC Material
The barcode scanner of the CoaguChek Pro II meter
can be used to scan barcodes from QC material vials.
This function can be set to
Optional
or
Scan Only
.
When set to
Scan Only
, the operator will not be able to
manually select a QC lot number from a list.
QC Range
Liquid quality control ranges can be customized to
comply with local guidelines. The QC Range function
enables you to narrow the default range for the test
parameter.
r
To set a QC range
3 Meter Setup
• 3
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......