87
CoaguChek
®
Pro II · Operator’s Manual
1
From the
Main Menu
, touch
Setup
to open the
meter settings.
2
From the
Setup Menu
, touch
QC Settings
.
3
From the
QC Settings
menu, touch
QC Range
.
4
From the
QC Range
menu, touch the button of the
test parameter you wish to set. You may select
from the following options:
•
Default Range
(Range provided in the code
chip is selected and displayed. No target value
is displayed.)
•
Custom Range
(The allowed percentage
deviation from target value can now be
customized. Additionally, you can choose
between displaying the target value along with
a control test result or not displaying the target
value with the control test result.)
5
Touch the button with the setting of choice. Your
selection is now highlighted.
•
If you selected
Default Range
, touch
to
save this setting, or touch
to exit this menu
without saving any changes.
•
If you selected
Custom Range
, touch
to
proceed with corresponding settings, or touch
to exit this menu without saving any
changes. The display automatically returns to
the previous screen.
3 Meter Setup
• 3
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......