89
CoaguChek
®
Pro II · Operator’s Manual
If the
Custom Range
and
Target Value
are set to
On
, they appear in a line below the control test
result in the QC test and QC memory screens.
The target value (2.9) appears
in front of
the
custom range.
If the
Display target value
is set to
Off
, only the
Custom Range
is displayed below the control test
result.
QC (quality control) Lockout
QC Lockout
ensures that the optional quality control
tests are run on a regular basis. This is independent of
the operator, however.
This option should be used if
you work without operator lists
.
QC Lockout
is disabled by default. For routine
confirmation of system reliability it is recommended
to enable this feature.
It is also possible to enable
Operator Lockouts
and
QC
Lockouts
in parallel. In addition to defining time
intervals or numbers,
QC Lockout
can also be used to
specify that a quality control test must be carried out
when a new test strip lot is used (
New Code
).
3 Meter Setup
• 3
Summary of Contents for CoaguChek Pro II
Page 1: ...CoaguChek Pro II Operator s Manual...
Page 15: ...15 CoaguChek Pro II Operator s Manual u Related topics 3 4 Options setup 62 1 Introduction 1...
Page 44: ...44 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 122: ...122 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 140: ...140 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 146: ...146 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 168: ...168 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 184: ...184 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 190: ...190 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 204: ...204 CoaguChek Pro II Operator s Manual Page intentionally left blank...
Page 205: ......