16
Rockwell Automation Publication 1441-UM002A-EN-P - May 2011
Chapter 2
Bump Test Extension Module
Set Up Bump Test
Measurements
Follow these steps to configure the parameters for the measurement.
1.
Connect the transducer to the Dynamix 2500 data collector.
2.
Attach the transducer to the machine case or structure.
3.
Select Bump Test on the Main Menu and press Enter.
The Bump Test screen appears.
•
Setup displays the Bump Test setup parameters and begins data
collection.
•
Recall lets you recall previously saved Bump Test settings to perform
another Bump Test using the same settings.
•
Review Data lets you review saved Bump Test data.
4.
Select Setup and press Enter to set up the Bump Test parameters and begin
to collect data.
Save – Saves the current bump
test settings. Enter a filename
to identify the bump test
settings.
Start – Take bump test data
using the current settings.
Back – Returns you to the Bump
Test menu.