12
E10037-180424
PROBEX
Make sure the conversion factors associated to this instrument have been entered. Refer
to the calibration certificate. For changing these, select ‘Calibration / Cal. Type / Review
Factory Calibration’ and select ‘Change factory calibration’, make the appropriate changes
and select ‘Save new factory calibration’ at the bottom of the screen. Finally select this
instrument.
Set the date and time of the Tablet if necessary.
Go in
‘Quick Readings’ and press the green icon for reading the sensors.
Two types of calibrations can be done, one with the probe unconfined (‘Pressure Loss
Calibration’) and one with the probe confined (‘Volume Loss Calibration’). See next section
for more details.
For configuring a ca
libration, go to ‘Calibration’ and :
Select the type of calibration
Give a calibration ID
Set the calibration timer