3-20
When “e” or “E” is displayed, perform any of the following actions according
to the captured image condition.
- Retake
Capture an image again if the images are not clear.
- Correcting outlines
Correct and redraw the incorrect images. Refer to “3.4.1 Editing
outlines
” for how to edit the
outlines.
- Selecting cell
Exclude cells with outlines extracted incorrectly from the target of
analysis. Refer to “3.4.1 c) Selecting cells” for how to select c
ells.
- Changing analysis method
Perform analysis with the L-count method or Core method. Refer
to “3.4.2 L
-
count method” and “3.4.3 Core method” for the
application of the L-count method and Core method.
The reliability mark does not appear when the following analytical
calculations are conducted.
- When an outline is edited
- When a cell is selected
- Analysis using the L-count method
- Analysis using the Core method
It can be set on the System Setup screen whether to show the reliability mark
or not.
Refer to “3.7.3 Application”
for setting method. The initial setting is OFF.
● Checking
Dark Area analysis
Touch the “Dark Area” button (1) on the single eye analysis screen (Fig. 3).
The Dark Area analysis scre
en (Fig. 4) appears. Touch the “Dark
Area
”
button (2) on the Dark Area analysis screen (Fig. 3) to return to the single
eye analysis screen.
(Fig. 3)
(Fig. 4)
(1)
(2)
Summary of Contents for REM 4000
Page 2: ......
Page 26: ...2 12 This page is intentionally left blank...
Page 33: ...3 7 Fig 1 Fig 2 2 1 3 4...
Page 82: ...3 56 This page is intentionally left blank...
Page 94: ...6 2 This page is intentionally left blank...
Page 101: ......