3-6
- Entry method 5: DICOM worklist inquiry
Setting
“
DICOM
”
to the system setup item
“
Auto Inquiry
”
, enables
DICOM worklist inquiries.
By pressing the
“
DICOM Inquiry
”
button (10) after entering the ID on
the
“
Patient Information
”
screen (Fig. 5), the patient information read
from the DICOM worklist in the personal computer appears.
(Fig. 5)
3)
Touch the “OK” button on the Patient Information screen (Fig. 2),
designate the patient information, and return to the previous screen.
-
Touch the “Cancel”
button (5) to discard the edited Information and
return to the previous screen.
3.2.5 Patient's eye height adjustment
■
When moving the head and/or chin rest, pay attention to the position of the
patient's face, hands, and fingers. Otherwise, injury to the hands and fingers
may result from the movement of the head and chin rest.
■
The chin rest paper is provided to keep the chin rest clean. For patient comfort,
use this chin rest paper, too.
■
Peel off the top sheet of chin rest paper and clean the forehead pad with a
clean cloth before capturing images for the next patient. Clean the forehead
pad with a cloth dampened with alcohol as needed.
1) Have the patient place their chin on the chin rest (1). Adjust the chin rest
height so that the height of the corner of the eye is aligned with the eye
level mark (2). Press the chin rest button (3) at the front of the main unit
(Fig. 2) to move the chin rest up and down.
2) When the patient's eye height is determined, lightly press the patient's
forehead against the forehead pad (4) to secure the patient's position.
(10)