3.4 Analysis
3.4.1 Auto Analysis
Auto Analysis is to calculate values such as CD (cell density) by automatically extracting
outlines of cells on the captured image
■
Check that the outlines of cells are correctly detected on the captured image.
The outlines cannot be extracted on some images.
■
Perform any of the following operations if outlines are not extracted correctly.
-
Retake
Capture an image again if the images are not clear.
-
Correcting trace line
Correct and redraw the incorrect images.
Refer to “3.4.1 b) Editing outlines” for how to edit the outlines.
-
Selecting cells
Exclude cells with outlines extracted incorrectly from the target of analysis.
Refer to “3.4.1 c) Selecting cells” for how to select cells.
-
Changing analysis method
Perform analysis with the L-count method or Core method. Refer to “3.4.2 L-
count method” and “3.4.3 Core method” for the operation of the L-count
method and Core method.
a) How to view analysis data
● R/L eye
capture screen
When touching the “Analyze” button on the capture screen or image selection
screen, the captured image is analyzed automatically and the R/L eye
analysis screen (Fig. 1) appears. In Quick mode, the image selection screen
appears when images of both eyes have been captured.
(Fig. 1)
(1)