SLIT LAMP RSL 4300/4500 by RODENSTOCK
SLIT LAMP RSL 4300/4500 by RODENSTOCK
RODENSTOCK SLIT LAMP
8. Eyepiece assembly securing knob
9. Yellow fi lter knob (up = out)
10. Illumination prism
11. Slit offset by rotating prism housing
12. Filter control wheel
13. Slit rotation knob
14. Slit width control knob
15. Aperture control wheel
16. Test bar & tonometer plate hole and cover
17. Illumination arm locking knob
18. Microscope arm locking knob
19. Diffuser
20. Joystick base locking knob
21. Joystick control (X Y Z movement)
22. Illumination control rheostat
23. Axle
24. Runner covers
25. Z type tonometer mounting hole
26. Yellow fi lter knob (up = out)
27. Lock for securing magnifi cation body
28. Magnifi cation change drum
29. Eyepieces – adjustable for PD and dioptre correction
30. Breath shield securing knob
21
22
23
24
25
28
27
29
30
26
20
3. CLEANING AND DISINFECTION INSTRUCTIONS
Before any cleaning of the instrument or the base unit, ensure the power
lead is disconnected.
Only manual non-immersion cleaning as described should be used for this instrument. Do not
autoclave or immerse in cleaning fl uids. Always disconnect power supply from source before
cleaning.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non- shedding cloth.
4. Safely dispose of used cleaning materials.
4. NAMES OF CONTROLS AND COMPONENTS
Headrest Assembly
1. Fixation light cable
2. Forehead rest band
3. Patient’s eye height marker
4. Fixation light
5. Chinrest
6. Chinrest height adjuster
7. Patient grab handles