7.2  ELECTROMAGNETIC IMMUNITY

Guidance and manufacturer’s declaration – electromagnetic immunity

The Rodenstock 

Digital Slit Lamp 

is intended for use in the electromagnetic environment 

specified below. The customer or user of the Digital Slit Lamp should ensure that it is used in 

such an environment

Immunity test IEC 60601 Test

level

Compliance 

level

Electromagnetic 

environment – guidance

Electrostatic 

discharge (ESD).
IEC 61000-4-2

± 8 kV contact
± 15 kV air

± 8 kV contact
± 15 kV air

Floors should be wood, concrete 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%.

Electrical fast 

transient/burst. IEC 

61000-4-4

± 2 kV for power 

supply lines
± 1 kV for power  

supply lines

N/A 

N/A 

Mains power quality should be 

that of a typical professional 

healthcare facility environment.

Surge.
IEC 61000-4-5

± 1 kV line(s) to 

line(s)
± 2 kV line(s) for 

input/output line(s)

N/A 

N/A 

Mains power quality should be 

that of a typical professional 

healthcare facility environment.

Voltage dips, short 

interruptions and 

voltage variations 

on power supply 

input lines.
IEC 61000-4-11

U

T

 = 0% 0.5 cycle

(0, 45, 90, 135, 180,
225, 270, 315°)

U

T

 = 0%; 1 cycle

U

T

 = 70%;

25/30 cycles (@ 0°)

U

T

 = 0%;

250/300 cycle

U

T

 = 0% 0.5 cycle

(0, 45, 90, 135, 180,
225, 270, 315°)

U

T

 = 0%; 1 cycle

U

T

 = 70%;

25/30 cycles (@ 0°)

U

T

 = 0%;

250/300 cycle

Mains power quality should be 

that of a typical professional 

healthcare facility environment. If 

the user of the Digital Slit Lamp 

requires continued operations 

during power mains interruptions, 

it is recommended that the Digital 

Slit Lamp be powered from an 

uninterruptible power supply.

Power frequency 

(50/60 Hz)
magnetic field. IEC 

61000-4-8

30 A/m

30 A/m

Power frequency magnetic fields 

should be at a level characteristic 

of a typical location in a typical 

professional healthcare facility 

environment.

Note: 

U

T

 is the a. c. mains voltage prior to application of the test level.

Immunity 

test

IEC 60601 

Test level

Compliance 

level

Electromagnetic environment 

– guidance

Portable and mobile RF communications equipment 

should be used no closer to any part of the Digital 

Slit Lamp, including cables, than the recommended 

separation distances calculated from the equation 

applicable to the frequency of the transmitter.

Recommended  separation distance

Conducted 

RF IEC 

61000-4-6

6 Vrms 150kHz 

to 80MHz

6 V

d = 1.2 √ p

Radiated 

RF IEC 

61000-4-3

10 V/m 80MHz 

to 2.7GHz

10 V/m

d = 1.2 √ p 80MHz to 800 MHz  

d = 2.3 √ p 800MHz to 2.7GHz

Where p is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separation 

distance in metres (m).
Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey¹, 

should be less than the compliance level in each 

frequency range.²

Interference may occur in the vicinity of 

equipment marked with this symbol.

Note: At 80MHz and 800MHz, the higher frequency range applies. These guide lines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
1 Field strengths from fixed transmitters, such as base stations ( cellular / cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 

accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 

should be considered. If the measured field strength in the location in which the Rodenstock Slit Lamp is used 

exceeds the applicable RF compliance level above, the Rodenstock Slit Lamp should be observed to verify normal 

operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientating or 

relocating the Rodenstock Slit Lamp.
2 Over the frequency range 150kHz to 80 MHz, field strengths should be less than 10 V/m.

7.3  RECOMMENDED SAFE DISTANCES

Recommended separation distances between portable and mobile RF 

communications equipment and the Digital Slitlamp

The Digital Slit Lamp is intended for the use in an electromagnetic environment in which radiated 

RF disturbances are controlled. The customer or the user of the Digital Slit Lamp can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile 

RF communications equipment (transmitters) and the Digital Slit Lamp as recommended below, 

according to the maximum output power of the communications equipment.

Summary of Contents for RSL 4500digital

Page 1: ...RSL 5000digital RSL 4500digital SLIT LAMP INSTRUCTIONS FOR USE Digital IFU Supplement ...

Page 2: ...OCK RSL 5000 SLIT LAMP 9 4 2 RODENSTOCK RSL 4500 SLIT LAMP 9 5 CONTROLS 10 5 1 BACKGROUND ILLUMINATION ADJUSTMENT 10 5 2 EXPOSURE AND REVIEW BUTTONS 10 5 3 CAPTURE BUTTON 10 5 4 KEYBOARD SHORTCUTS 10 6 WARRANTY 11 7 SPECIFICATIONS AND ELECTRICAL RATINGS 11 7 1 ELECTROMAGNETIC EMISSIONS 11 7 2 ELECTROMAGNETIC IMMUNITY 12 7 3 RECOMMENDED SAFE DISTANCES 13 8 TECHNICAL SPECIFICATIONS 14 9 DIGITAL ACCE...

Page 3: ... examination of the anterior eye segment from the cornea epithelium to the posterior capsule It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment Brief description of the instrument This Rodenstock Slit Lamp can either be mounted onto a custom table top supplied by Rodenstock or can be mounted on a third parties table top refr...

Page 4: ... is used only in the configuration as supplied Should the instrument suffer shocks for example should it accidentally fall and the optical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair Do not touch accessible connectors and the patient simultaneously The owner of the instrument is responsible for training personnel in its ...

Page 5: ...DISINFECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do not autoclave or immerse in cleaning fluids Always disconnect power supply from source before cleaning 1 Wipe the external surface with a clean absorbent non shedding cloth dampened with de ionis...

Page 6: ... ADJUSTMENT long end short end RSL 5000 SLIT LAMP RSL 4500 SLIT LAMP 6 Connect the shorter USB cable 3020 P 7028 to the socket on the underside of the DCA and to the appropriate socket on the USB hub located on the Slit Lamp base Ensure that there is plenty of slack at the USB hub end to allow Slit Arm rotation 7 Route the cable in the groove in the Slit Lamp arm and replace the cover plate 8 Conn...

Page 7: ...is instrument Any servicing or repairs should only be carried out by Rodenstock Instruments or by suitably trained and authorised distributors Service manuals will be available to authorised Rodenstock service centres and Rodenstock trained service personnel 7 SPECIFICATIONS AND ELECTRICAL RATINGS The Rodenstock Digital Slit Lamp is a medical electrical instrument The instrument requires special c...

Page 8: ...t of the Digital Slit Lamp including cables than the recommended separation distances calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF IEC 61000 4 6 6 Vrms 150kHz to 80MHz 6 V d 1 2 p Radiated RF IEC 61000 4 3 10 V m 80MHz to 2 7GHz 10 V m d 1 2 p 80MHz to 800 MHz d 2 3 p 800MHz to 2 7GHz Where p is the maximum output power ra...

Page 9: ... system Microsoft Windows 10 Pro Windows 10 Pro Recommended screen resolution 1920 x 1200 pixels Software requirements CE compliant or FDA cleared camera imaging software Weight packed approx Digital Slit Lamp complete 20 0Kg 75 x 54 x 45cm W x D x H Protection against ingress IPxO Class II ME equipment Insulation between mains parts and the functional earth provide at least two means of protectio...

Page 10: ...Handle EP79 80070 1 27mm A F ALLEN KEY EP39 80329 1m DSL to PSU Cable EP39 80321 5mm L Shaped Hexagonal Key 3020 P 7077 6mm L Shaped Hexagonal Key 3020 P 5221 Black Patient Handle 3020 P 5039 Cable Clip 3020 P 7054 Chin Rest Pins 406 EP39 80365 Digital Base to Computer Cable EP39 80364 DSL Accessory Kit Packaging EP39 80250 DSL Digital Background Illum Assy 3020 P 7014 DSL Digital Backlight Filter...

Page 11: ...he environmental impact of WEEE Waste Electrical Electronic Equipment and minimise the volume of WEEE entering landfills we encourage at product end of life that this equipment is recycled and reused If you need more information on the collection reuse and recycling then please contact B2B Compliance on 01691 676124 44 1691 676124 UK only Any serious incident that has occurred in relation to the d...

Page 12: ...800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 39800 Issue A Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Straße 21 90427 Nürnberg Germany Tel 49 9131 777 1 777 Fax 49 9131 777 1 20 Visiometrics S L Vinyals 131 08221 Terrassa Spain ...

Reviews: