background image

Common drain, blown-down and transfer lines   

24    Independent drain lines and blow-down lines should be 
provided. If this is not practicable, devices should be provided 
to ensure that only one autoclave blow-down valve can be in 
the open position at any one time, in order to safeguard 
against inadvertent pressurisation or scalding. Where the 
exhaust steam from one autoclave is re-used in another 
autoclave, it is essential to provide a valve in the transfer line, 
which is interlocked with the door locking mechanism of the 
autoclave to which the steam is being supplied.   

25    No person should be permitted to enter the confined 
space of an autoclave unless an assessment has been made 
of the risks to health and safety. This is particularly important 
when fumes or an inert atmosphere such as nitrogen may be 
present. Following the assessment a safe system of work 
should be drawn up and implemented. Persons entering an 
autoclave should use a key or tag to prevent the door being 
closed behind them.   

Autoclaves for sterilizing liquids in sealed glass 
containers  

26 

To reduce the possibility of sealed glass containers 

exploding when an autoclave sterilizer door is opened, a 
temperature sensing device should be provided which 
controls a door interlock arranged to prevent the door opening 
until the temperature in all the containers has fallen to less 
than 80

0

C

The sensing device may be inserted into a 

simulator, designed so that its temperature on cooling is 
never less than that of any part of the load. Where it is not 
reasonably practicable to provide a temperature sensitive 
interlock, a time activated interlock may be used set to ensure 
an adequate cooling period. Whatever method is used, the 
controls should be tamper-proof and should only be set or 
adjusted by authorised persons.   

INTERLOCKING 

27 

In the context of autoclave safety, any interlock should 

ensure that the door cannot be opened when potentially 
dangerous conditions exist within the autoclave and that a 
cycle cannot be initiated until the door is completely closed, 
the securing mechanism is fully and correctly engaged and 
the chamber sealed.   

 28    Some of the situations in which interlocks should be 
provided have been described in paras 18 to 26.   

 
29   

Because an operator relies on the safety afforded by 

interlocks it is essential that all interlocks should fail to safety. 
It is also essential that the interlock devices are sufficiently 
robust and securely attached to the autoclave prevent them 
being distorted, overridden, or defeated during normal 
operation. They should function reliably, be properly 
maintained and be kept correctly adjusted, strictly in 
accordance with the manufacturers instructions. This is 
particularly important where safety is dependent on limit 
switches. If these are out of 

 

 
 
 
 
 
 
 
 
 

Adjustment they could allow premature operation of a 
following sequence with dangerous consequences. Useful 
guidance on the principles of application of interlocking 
devices is given in BS5304: 

Safety of machinery.

 

 30   

A properly trained and authorised person should 

check at regular intervals that the interlocks function correctly 
and cannot be easily overridden or defeated; this should be 
done at least daily if the autoclave is in constant use. 

TRAINING 

31   

Operators of autoclaves should be fully instructed and 

properly trained in correct operating procedures and the 
purpose and function of controls and safety devices. They 
and their managers should be aware of the dangers of 
by-passing or interfering with safety devices or of other 
malpractices. Manufacturers' instructions and operating 
procedures should be readily available to operators and 
should always be adhered to and any malfunction or faulty 
equipment should be reported without delay to an appointed 
superior who will ensure that appropriate remedial action is 
taken.  

EXAMINATION AND MAINTENANCE 

Initial testing and inspection 

32   

An initial pressure test in accordance with the design 

standard should be carried out on each autoclave to 
demonstrate its integrity as far as it is possible with a test of 
this nature. This test should be carried out in the presence of 
a competent person, who may be the representative of an 
independent engineering inspection organisation. All 
interlocks and safety devices should be assessed and 
inspected for correctness of operation before the autoclave is 
taken into use. A report of the test and inspection should be 
made available to the user and retained by him.   

Note: 

Pneumatic testing is not recommended, as it is 

hazardous and requires special safety precautions. 
Advice is contained in Guidance Note GS4 - 

Safety in 

pressure testing 

and in BS 5500.   

33   

In addition to any information which may be required 

by the relevant legislation, the following information should be 
permanently marked on the vessel:-   

(a) 

The name of the manufacturer or his identifying 
mark.  

 
(b) 

The serial number 

 
 
(c) 

The design pressure 

 
 
(d) 

The design temperature 

 
 
(e) 

The hydraulic test pressure 

 
 
(f) 

The date of hydraulic test 

 
 
(g) 

Date of manufacture and manufacturing standard. 

 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for GEMINI-MP25 with SmartCard

Page 1: ...Operators Manual GEMINI MP25 with SmartCard Serial number 2140 Rodwell Scientific Instruments Bentalls Basildon Essex England Tel 44 0 1268 286646 Fax 44 0 1268 287799 Manual issue 14 12 10...

Page 2: ...ECHNICAL INFORMATION 7 RISK ASSESSMENT AND SAFETY 8 USER REPONSIBILITES 9 A GUIDE TO THE PLANNING AND IMPLEMENTATION OF YOUR HEALTH AND SAFETY CONTROL MEASURES 10 DISPOSAL CARING FOR THE ENVIRONMENT 1...

Page 3: ...a means of breaking the pressure seal with the door restrained The development of the Rodwell GEMINI has resulted in a unit that is both easy to use and service SAFETY NOTE We recommend that a copy o...

Page 4: ...ard water areas the supply should be treated to avoid chamber element and pipe scaling Distilled water water softening units that use salts reverse osmosis systems or de ionised water supplies are not...

Page 5: ...packing materials are removed from inside the autoclave chamber Fill the chamber with water until display stops displaying Low water message Close the door and tighten the bolt tip of bolt must regist...

Page 6: ...nt program number sterilise temperature and period If the power was interrupted during a program cycle then if it is possible the machine will continue with the program if this is not possible for exa...

Page 7: ...ressing the up or down buttons the program details will be displayed or will be blank if a program number has not been set up If the machine has the option for Instrument programs Where the chamber is...

Page 8: ...r open is pushed if you do not wish to remove the load at this time but want to stop the sounder operating it is also possible to push the stop button to return the unit to Select prog Mode without re...

Page 9: ...they are displayed in minutes except for the sterilise time which is in hours and minutes temperature is always in degrees centigrade Note that some functions described may not operate on your machin...

Page 10: ...top button will cause the machine to abort the current cycle and to immediately start cooling Standby If set to on the water temperature will be maintained at a preset temperature to speed up the heat...

Page 11: ...has occurred take any necessary precautions as the load my not be sterile NOTE you will not be able to reset a lockout condition until the temperature has fallen below 80 C USB Memory This option will...

Page 12: ...to operate the memory stick must be inserted into the USB socket on the control panel Do not remove the memory stick while a cycle is in progress as the system cannot then log the cycle and data may b...

Page 13: ...n suffix MSL is applied to all log names Using the logger display program Please refer to the separate manual for information on using this program A PDF file on the memory stick explains fully how to...

Page 14: ...low water if filling is required Close the door and tighten the bolts until you hear a beep Bottom bolt first Choose the required program cycle if the correct program is not already set then change th...

Page 15: ...epeating the above sequence To re start the cycle simply press the start button once the cause of the fault has been rectified Pressure switch The safety door interlock pressure switch failed to opera...

Page 16: ...thermal lock solenoid does not pull in then the door cannot be opened Turn the bottom lock bolt a little in each direction to try and free it Water level low Out of water Fill with water and restart...

Page 17: ...art Door not closed Tighten door bolts DO NOT FORCE Water level Low Fill with water Water level is not Water too pure detected by probes Setting too high Reset temperature 136 degrees C max Fails to r...

Page 18: ...rinter mechanism press and hold the paper feed button until the mechanism takes the paper and it starts to appear from the front of the printer 4 Ensure that the paper follows the correct path as per...

Page 19: ...Correct paper path Main door release catch Printer release catch PUSH Figure 1 Figure 2 Figure 3 Figure 4 Ribbon above paper Feed paper between ribbon and cartridge...

Page 20: ...fitted 120 5 C 1 0bar 134 C 2 0bar 2 Check that indicating display and alarms function correctly 3 Check pipe connections for signs of leaks 4 Lightly lubricate the door bolt with high melting point c...

Page 21: ...d in the form of a Preventative Maintenance Contract along with Rodwell or NAMAS Calibration Services For details of Service Calibration Contracts contact Rodwell Scientific Instruments Service Depart...

Page 22: ...FE DESIGN OPERATING LIMITS Maximum Working Temp 138 C Minimum Working Temp 7 C Maximum Hydraulic Pressure 4 14 Bar Maximum Working Pressure 2 26 Bar Maximum Working Vacuum Minus1 Bar SODIUM CHLORIDE W...

Page 23: ...through loading and unloading Use Loading Hoist or Loading Trolley Burns and Scolds from opening the autoclave before cycle complete Locks fitted However care should always be taken Damage to Load Pro...

Page 24: ...odium chlorides from the load or from the water used in filling the vessel the operator carries out routine maintenance checks as stated in this manual the operator must is fully trained before using...

Page 25: ...y Records Earth Continuity Test on all metal panels recorded This is a Safety Critical Operation Insulation Type Class1 for Insulation Resistance Test Safety Critical Components adjusted and maintaine...

Page 26: ...y If the Autoclave is to be permanently demolished and or scrapped divide the material to be disposed of according to type and composition as follows 1 Iron or ferrous materials composed of metal alon...

Page 27: ...rent legislation s and guide lines relevant to the country in which the equipment is to be used are obtained and adhered to for the safety of personnel and the safe operation of the equipment Whilst e...

Page 28: ...rts for 2 years 2 off 312 0009 Door Seal 6 off 315 0201 Heater Element 277V 4 off 309 5038 Heater Glands 6 off 836 0180 PT100 Black Probes 2 off 820 0067 Bolt Safety Switch 2 off 309 0763 Thermal Lock...

Page 29: ...piry date Comments Signature Service 3 Date Machine cycle count Engineer s Name Company Name Rodwell certification No Certification expiry date Comments Signature Service 4 Date Machine cycle count En...

Page 30: ...neer s Name Company Name Rodwell certification No Certification expiry date Comments Signature Service 16 Date Machine cycle count Engineer s Name Company Name Rodwell certification No Certification e...

Page 31: ...ick opening doors locking mechanisms may be manually or power operated In the following text the term door should be taken as including the lid of a vertical autoclave DANGERS 7 Autoclaves give rise t...

Page 32: ...s SAFEGUARDS Design 14 All parts of an autoclave its foundations and supporting structures should be properly designed of good construction sound material and adequate strength The requirements for ac...

Page 33: ...inated That may be achieved by using an interlocking arrangement incorporating a temperature sensing device located in that part of the vessel which is expected to remain at the highest temperature at...

Page 34: ...tly robust and securely attached to the autoclave prevent them being distorted overridden or defeated during normal operation They should function reliably be properly maintained and be kept correctly...

Page 35: ...affected by a work activity 37 The Factories Act 1961 Section 30 may apply see HSE Guidance Note GS5 Entry into confined spaces 38 The Pressure Systems and Transportable Gas Container Regulations 1989...

Reviews: