If any other messages appear then press the door open button and carry out the required
operation before pushing the start button again.
Pressing the door open or the stop button will stop the process cycle until the temperature
reaches 80
C. Once the temperature reaches this figure the process cannot be stopped other
than by use of the stop button.
The machine will work through the various stages of sterilising, displaying appropriate
information on the screen.
Cycle complete
Once the cycle is completed and the machine has cooled to below 80
C it will display
Complete
and the sounder will beep every minute. If the cycle completed without any
problems the machine will display Load is sterile, but if any problems occurred the message
will show Load not sterile. If the load is not sterile take any necessary precautions to avoid
contamination.
Press the door open button to release the door. The door light will come on and the display
will show
door is released
and then
Door is unlocked
If the display shows
door lock
jammed
then turn the top bolt slightly clockwise or anticlockwise to release the bolt, now
push the door open button again. If the display shows
pressure locked
then move the
bottom bolt slightly clockwise or anticlockwise to release the pressure lock.
Once the display indicates
door is released
undo the bolt and open the door.
Note that the display returns to the
Select prog
when the door open is pushed, if you do not
wish to remove the load at this time but want to stop the sounder operating it is also possible
to push the stop button to return the unit to
Select prog
Mode without releasing the door
Remove the load.
The machine is now ready for further use.
Lockout
If during the sterilising process the system should generate a fault the machine will go into
Lockout mode. The screen will indicate what caused the problem and the sounder will beep
every minute until the condition is reset. The sterilising cycle will be aborted and the
machine cooled down. Once the machine has cooled and the condition reset it will be
possible to open appropriate door but consider any health risks involved as the load may not
have been correctly sterilised. If the machine consistently locks out then consult your service
agent as there may be a problem with the machine
There are two other buttons on the control panel
Delayed start
If you wish to make use of cheap rate electricity supplies or to delay the start of sterilising so
that the process is completed for the morning, a delayed start function is incorporated. Load
the machine as described above, close the door and select the required program. Do not press
start.
Press the delayed start button. The indicator will light and the screen will display the start
time with the hours flashing. Press the Increase or decrease buttons to set the hours to the
time that you wish the machine to start. (Use 24 hour format.) Press the right shift button to
move to the minutes and set as required.
Ensure that the correct program has been selected then press start.