SMIQ
RF Frequency
1125.5555.03
E-8
2.45
2.4 RF
Frequency
The RF frequency can be set directly using the [FREQ] key (cf. Section 2.2.6) or by accessing menu
FREQUENCY.
The frequency of the RF output signal is entered/indicated under FREQUENCY in the FREQUENCY
menu.
The input value of frequency settings opened by means of the [FREQ] key and indicated in the header
line considers the offset in calculation (cf. next Section). This offers the possibility of entering the desired
output frequency of possibly series-connected instruments such as mixers in the menu.
Note:
Further settings:
Frequency sweep
Menu SWEEP
LF frequency
Menu ANALOG MOD
Menu LF-OUTPUT
Int./ext. reference frequency
Menu UTILITIES-REF OSC
Phase of the output signal
Menu UTILITIES-PHASE
Menu selection:
FREQUENCY
FREQUENCY
OFFSET
KNOB STEP USER
KNOB STEP
EXCLUDE FROM RCL
100. 000 000 0 MHz
0.0
Hz
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
1. 000 000 0 MHz
DECIMAL
USER
ON
OFF
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-16
Menu FREQUENCY (preset setting)
FREQUENCY
Input value of the RF frequency at the RF output connector.
IEC/IEEE-bus command
SOUR:FREQ 100E6
OFFSET
Input value of the frequency offset, e.g., of a series-connected mixer (cf.
Section "Frequency Offset"). The status line indicates FREQ-OFFST.
IEC/IEEE-bus command
SOUR:FREQ:OFFS 0
KNOB STEP USER
Input value of the step width for frequency variation using the rotary knob. The
RF frequency is varied in the step width entered if KNOB STEP is set to USER.
IEC/IEEE-bus command
SOUR:FREQ:STEP 1MHz
KNOB STEP
DECIMAL:
Variation step width corresponding to the position of the digit
cursor.
USER:
"User Defined", variation step width as entered under KNOB
STEP USER .