Digital Standard DECT
SMIQ
1125.5555.03
E-9
2.328
SLOT ATTENUATION
Input value for the level attenuation in dB of all active slots whose SLOT
LEVEL was set to ATTEN. The slots whose level is to be attenuated are
defined in the menu SELECT SLOT.
IEC/IEEE-bus command
:SOUR:DECT:SLOT:ATT 40 DB
TIMING
ADJUSTMENT
SIMULATION
Input value for the number of bits used for simulating "Timing Adjust" in DECT
instruments. Each 35
th
frame is extended (positive input values) or shortened
(negative values) by the number of bits set.
IEC/IEEE-bus command
:SOUR:DECT:SIM:TADJ -1
JITTER SIMULATION
Input value for the number of bits used for jitter simulation. A jitter is simulated
by advancing the selected number of bits for even-numbered frames in time
whereas ordinary timing is generated for odd-numbered frames.
IEC/IEEE-bus command
:SOUR:DECT:SIM:JITT 1
PREAMBLE
Selection of the preamble type for full and double slot.
NORMAL
The preamble field has a length of 16 bits.
PROLONGED The preamble field has a length of 32 bits.
IEC/IEEE-bus command
:SOUR:DECT:PRE NORM
SAVE/RCL FRAME...
Opens a window for saving and loading a frame configuration. Loading a
frame affects all parameters that can be set under SELECT SLOT.
FREQ
100. 000 000 0
MHz
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
DECT
PHS
IS95
NADC
PDC
GSM
DECT
TRIGGER. ...
CLOCK...
POWER RA MP C>
SLOT ATT ENUA
TIMING A DJUS
JITTER S IMUL
PREAMBLE
SAVE/RCL FRA
GET PREDEFINED FRAME...
RECALL FRAME...
SAVE FRAME...
DELETE FRAME...
LEVEL
- 30.0
dBm
PEP
- 30.0
dBm
FULL
0
1
Fig. 2-210
Menu DIGITAL STD - DECT - SAVE/RCL FRAME, SMIQ equipped with Modulation Coder
SMIQB20 and Data Generator SMIQB11