4.2.2 Status Information and Messages
The status information and messages are displayed in the header section of the screen.
The messages differ with respect to their importance (errors, warnings, info) and the time
of their appearance (brief and permanent messages). They require different treatment
by the user. Further information on all messages can be called in the info window (see
Refer to "Error messages" in the Operating Manual for an overview of all status informa-
tion and messages as well as instructions for error elimination.
4.2.2.1
Status Information
The status information gives the user an overview of the main operating states and set-
tings of the instrument. The states are indicated for information only and do not neces-
sitate any action by the user.
Status information is displayed between the frequency and level fields, at the left of the
info line or in the info line itself.
On two-path instruments, all states that can occur independently in the two paths are
displayed separately for each path. The associated path is indicated in the info line.
4.2.2.2
Messages
Messages indicate errors in the instrument. They are displayed in the info line in different
colors depending on their importance and display duration. Errors (e.g. no calibration
data) are displayed in red, information (e.g. file not found) and warnings in black. Warn-
ings indicate less significant errors (e.g. the instrument operates outside specified data).
4.2.2.3
Volatile messages
Brief messages report automatic settings in the instrument (e.g. switching off of incom-
patible types of modulation) or on illegal entries that are not accepted by the instrument
(e.g. range violations). They are displayed in the info line on a yellow background. They
are displayed on top of status information or permanent messages.
Volatile messages do not normally demand user actions and disappear automatically
after a brief period of time. They are stored in the history, however.
These messages can be read from remote using
SYST:ERR?
or
SYST:ERR:ALL?
.
Display
Summary of Contents for 1141.2005.02
Page 27: ......
Page 74: ...Preparing for Use R S SMU200A 57 Quick Start Guide 1007 9822 62 13 LXI Configuration ...
Page 75: ......
Page 95: ......
Page 135: ......
Page 141: ......
Page 144: ...Index R S SMU200A 127 Quick Start Guide 1007 9822 62 13 Winbar 89 Windows XP 33 access 34 ...