Preparing for Use
R&S
®
SMU200A
50
Quick Start Guide 1007.9822.62 ─ 13
3. To terminate the connection on the external Windows PC, close the VNC Viewer
program.
The connection is terminated. The color of the VNC icon in the status bar of the
instrument changes.
A message on the external PC indicates the disconnection.
Disabling Remote Access via Ultr@VNC
The VNC connection can be disabled by removing the program on the instrument or by
deactivating the VNC Server service.
1. Remove the VNC program
a) Press the "Windows" key to access the operating system and open the "Add or
Remove Programs" by selecting "Start > Settings > Control Panel > Add or
Remove Programs".
b) Remove the VNC program.
2. Deactivate the VNC Server service
a) Press the "Windows" key to access the operating system and open the "Ser-
vices" by selecting "Start > Settings > Control Panel > Services".
b) Deactivate the VNC Server service.
The connection is disabled, the VNC icon disappears from the task bar of the instru-
ment.
2.8 Using Virus-Protection Software
Virus-protection software Symantec Antivirus 10.1.7.7000 has been successfully tested
for compatibility with the instrument software on R&S Signal Generator.
However, Symantec Antivirus may affect the behaviour of the instrument in terms of set-
tling time or stability. Symantec Antivirus 10.1.7.7000 has been subjected to detailed tests
with R&S Signal Generator and a page file of 1024 MB. The system remains very reliable.
However, rare impact on the settling time has been experienced ( slight degradations of
settling time in 2% of samples out of specification).
Preparing the Installation of Symantec Antivirus
The installation is performed in the Windows XP operating system.
Accessing Operating System
The operating system can only be accessed if an external keyboard and/or a mouse is
connected. Windows XP provides an on-screen keyboard that can be operated using a
mouse, thus enabling operation by mouse only. An external keyboard is recommened for
extensive entries.
1. Press the "Windows" key to access the operating system and open the "System"
dialog by selecting "Start > Settings > Control Panel > System".
Using Virus-Protection Software
Summary of Contents for 1141.2005.02
Page 27: ......
Page 74: ...Preparing for Use R S SMU200A 57 Quick Start Guide 1007 9822 62 13 LXI Configuration ...
Page 75: ......
Page 95: ......
Page 135: ......
Page 141: ......
Page 144: ...Index R S SMU200A 127 Quick Start Guide 1007 9822 62 13 Winbar 89 Windows XP 33 access 34 ...