Preparing for Use
R&S
®
SMU200A
25
Quick Start Guide 1007.9822.62 ─ 13
Risk of instrument damage
Before switching on the instrument, make sure that the following conditions are met:
●
Instrument casing is closed and all fasteners are tightened.
●
All fan openings are unobstructed and the airflow perforations are unimpeded. The
minimum distance from the wall is 10 cm.
●
The instrument is dry and shows no sign of condensation.
●
The instrument is operated in the horizontal position on an even surface.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are correctly connected and are not overloaded.
Failure to meet these conditions may cause damage to the instrument or other devices
in the test setup.
Risk of electrostatic discharge
Protect the work area against electrostatic discharge to avoid damage to electronic com-
ponents in the modules. For details, refer to the safety instructions at the beginning of
this manual.
EMI impact on measurement results
Electromagnetic interference (EMI) can affect the measurement results. To avoid any
impact, make sure that the following conditions are met:
●
Use suitable double-shielded cables.
●
Do not use USB connecting cables exceeding 1 m in length.
●
Use only USB devices that remain within the permissible EMI limits.
●
Always terminate any connected IEC-bus cables with an instrument or controller.
Outputs and lines have to be terminated properly with 50 Ohms.
For digital interfaces only the cable R&S SMU-Z6 is permitted for connection. The asso-
ciated cable is available under part number 1415.0201.02.
2.3.1 Unpacking and Checking the Instrument
To remove the instrument from its packaging and check the equipment for completeness
proceed as follows:
1. Pull off the polyethylene protection pads from the instrument's rear feet and then
carefully remove the pads from the instrument handles at the front.
Putting into Operation
Summary of Contents for 1141.2005.02
Page 27: ......
Page 74: ...Preparing for Use R S SMU200A 57 Quick Start Guide 1007 9822 62 13 LXI Configuration ...
Page 75: ......
Page 95: ......
Page 135: ......
Page 141: ......
Page 144: ...Index R S SMU200A 127 Quick Start Guide 1007 9822 62 13 Winbar 89 Windows XP 33 access 34 ...