Preparing for Use
R&S
®
SMU200A
30
Quick Start Guide 1007.9822.62 ─ 13
●
Uninterrupted level settings are switched off
"Level Attenuator Mode" = AUTO
●
Internal level control "Level ALC" = AUTO
●
User correction "Level Ucor" = OFF
●
"LF output State" = Off
●
"Sweep State" = Off
●
"List mode State" = Off
Settings that are not affected by the PRESET key
●
Reference frequency settings ("Ref Oscillator" menu)
●
Power on settings ("Level/EMF" menu)
●
Network settings ("Setup" menu)
●
GPIB address ("Setup" menu)
●
Password and settings protected by passwords ("Setup" menu)
●
Start/Stop Gui Update ("Setup" menu)
●
Display and keyboard settings ("Setup" menu)
User-defined instrument states can be stored and called up in the "File" dialog.
2.3.5 Shutting Down the Instrument
To shut down the R&S SMU, proceed as described below.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting
the power cord, the instrument loses its current settings. Furthermore, program data may
be lost.
Press the ON/STANDBY key first to shut down the application properly.
1. Press the ON/STANDBY key to save the current setup, shut down the operating
system and set the instrument to standby state.
The yellow LED must be on.
2. To turn the power off:
a) For instruments with power switch, press the main power switch to position 0
(Off).
b) For instruments without power switch, pull the power cable at the rear of the
instrument.
Tip:
Putting into Operation
Summary of Contents for 1141.2005.02
Page 27: ......
Page 74: ...Preparing for Use R S SMU200A 57 Quick Start Guide 1007 9822 62 13 LXI Configuration ...
Page 75: ......
Page 95: ......
Page 135: ......
Page 141: ......
Page 144: ...Index R S SMU200A 127 Quick Start Guide 1007 9822 62 13 Winbar 89 Windows XP 33 access 34 ...