Remote-control commands
DVB-H
66
Operating Manual 1008.0970.12 ─ 16
Example:
BB:DVB:TRIG:OUTP2:PULS:DIV 2
sets the divider for the corresponding marker signal to the value
2.
BB:DVB:TRIG:OUTP2:MODE PULS
enables the pulsed marker signal.
BB:DVB:TRIG:OUTP:FREQ?
queries the resulting pulse frequency of the marker signal.
Response:
66 000
the resulting pulse frequency is 66 kHz.
Usage:
Query only
Manual operation:
See
4.1.6 Clock Settings
This section lists the remote control commands, necessary to configure the clock.
[:SOURce<hw>]:BB:DVB:CLOCk:MODE
............................................................................66
[:SOURce<hw>]:BB:DVB:CLOCk:MULTiplier
...................................................................... 66
[:SOURce<hw>]:BB:DVB:CLOCk:SOURce
.........................................................................67
[:SOURce<hw>]:BB:DVB:CLOCk:SYNChronization:EXECute
...............................................68
[:SOURce<hw>]:BB:DVB:CLOCk:SYNChronization:MODE
.................................................. 68
[:SOURce<hw>]:BB:DVB:CLOCk:MODE
<Mode>
The command enters the type of externally supplied clock (
BB:DVB:CLOCk:SOURce
EXTernal
). When
MSAM
is used, a multiple of the sample clock is supplied and the
clock is derived internally from it. The multiplier is entered with the com-
mand
:BB:DVB:CLOCk:MULTiplier
.
For two-path instruments, the only numerical suffix allowed for SOURce is 1, since the
external clock source is permanently allocated to path A.
Parameters:
<Mode>
SAMP | MSAMp
*RST:
SAMP
Example:
BB:DVB:CLOC:MODE MSAM
sets the type of externally supplied clock.
Manual operation:
See
[:SOURce<hw>]:BB:DVB:CLOCk:MULTiplier
<Multiplier>
The command specifies the multiplier for clock type "Multiple Sample"
(
:BB:DVB:CLOCk:MODE MSAMple
) in the case of an external clock source.
For two-path instruments, the only numerical suffix allowed for
SOURce
is 1, since the
external clock source is permanently allocated to path A.