Remote Control Commands
R&S
®
NGP800
109
User Manual 5601.5610.02 ─ 04
Suffix:
<IO>
.
1..8
Parameters:
<state>
1
Selected Digital /O line is enabled.
0
Selected Digital /O line is disabled.
*RST:
0
TRIGger[:ENABle]:GENeral
<arg0>
Sets or queries the enable state of the master on/off of Digital I/O trigger.
Parameters:
<master_state>
1
Master state of Digital I/O trigger is enabled.
0
Master state of Digital I/O trigger is disabled.
*RST:
0
TRIGger[:ENABle]:SELect:DIO<IO>
<arg0>
Sets or queries the enable state of the specified Digital I/O line.
Suffix:
<IO>
.
1..8
Parameters:
<state>
1
The specified Digital I/O line is enabled.
0
The specified Digital I/O line is disabled.
7.5
Configuration Commands
The following subsystems contain the commands for channel selection, voltage and
current settings for the instrument.
7.5.1
Channel Selection
The
INSTrument:Select
subsystem contains the commands for selecting the output
channels.
Each channel of the power supply is considered as separate "instrument", which is
required by the SCPI standard. Therefore, the SCPI commands use the
INSTRument
node to select a channel.
Configuration Commands
www.allice.de
Allice Messtechnik GmbH